Araştırma özeti
Aşağıdaki metin kaynak makalenin lisansı doğrulanmış orijinal özetidir. cc by
BackgroundParkinson's disease is a neurodegenerative disorder that is characterized by a degeneration of the dopaminergic system. Dopamine transporter (DAT) positron emission tomography (PET) imaging has emerged as a powerful and non-invasive method to quantify dopaminergic function in the living brain. The PET radioligand, [18F]FE-PE2I, a cocaine chemical derivative, has shown promising properties for in vivo PET imaging of DAT, including high affinity and selectivity for DAT, excellent brain permeability, and favorable metabolism. The aim of the current study was to scale up the production of [18F]FE-PE2I to fulfil the increasing clinical demand for this tracer.ResultsThus, a fully automated and GMP-compliant production procedure has been developed using a commercially available radiosynthesis module GE TRACERLab FX2 N. [18F]FE-PE2I was produced with a radiochemical yield of 39 ± 8% (n = 4, relative [18F]F- delivered to the module). The synthesis time was 70 min, and the molar activity was 925.3 ± 763 GBq/µmol (250 ± 20 Ci/µmol). The produced [18F]FE-PE2I was stable over 6 h at room temperature.ConclusionThe protocol reliably provides a sterile and pyrogen-free GMP-compliant product.
Bilimsel değerlendirme
Araştırma sorusu
Sorunun kapsamı orijinal çalışma üzerinden değerlendirilir.
Yöntem ve örneklem
Çalışma tasarımı ve örneklem için kaynak makaleyi kontrol edin.
Temel bulgular
Bu kayıt için editoryal olarak onaylanmış bulgu özeti henüz yayımlanmadı.
Sınırlılıklar
Sonuçların genellenebilirliği, yanlılık ve klinik anlamı ayrıca incelenmelidir.
Bibliyografik veriden yöntem, örneklem veya klinik sonuç çıkarımı yapılmaz.