When battery exhaustion lets the lame walk: a case report on the importance of long-term stimulator monitoring in deep brain stimulation.

BMC Neurol · 2015-07-19

DOI: 10.1186/s12883-015-0365-6PMID: 26187628

Open accessJournal Article; Case Reports; case-reportOriginal source ↗

Research overview

The text below is the source article’s original abstract under a verified reuse license. cc by

BackgroundDeep brain stimulation is increasingly used in the treatment of advanced Parkinson's disease. While its short-term effectiveness is well documented, there are only few reports on long-term outcomes, and the need to repeatedly reprogram the stimulator is seldom reported.Case presentationWe present a 74-year-old man with gait impairment, which had been mistaken for worsening of the disease and only remitted when the stimulator battery was exhausted indicating that the stimulator itself had been the cause.ConclusionThis case highlights the need to repeatedly monitor not only battery capacity, but also stimulator-related side-effects for an extended period after implantation and, if necessary, to refer to centres capable of systematically reprogramming the device.