SCIENTIFIC HUB

Follow the evidence.

SCIENTIFIC LITERATURE · MED

Validation of Blood Pressure Measurement Using a Smartwatch in Patients With Parkinson's Disease.

Front Neurol · 2021-06-29

DOI: 10.3389/fneur.2021.650929PMID: 34267718

Open accessbrief-report; Journal ArticleOriginal source ↗

Research overview

The text below is the source article’s original abstract under a verified reuse license. cc by

Objectives: We aimed to validate the accuracy of blood pressure (BP) measurement using a smartwatch in patients with Parkinson's disease (PD). Materials and Methods: We compared 168 pairs of BP (n = 56) measurements acquired by a smartwatch (SM-R850) with those measured by a sphygmomanometer (reference device). Results: Differences between the smartwatch BP and reference BP measurements were compared. The mean and standard deviation of the differences systolic BP (SBP) and diastolic BP (DBP), measured by smartwatch and reference device, fulfilled both criterion 1 (0.4 ± 4.6 and 1.1 ± 4.5 mm Hg for DBP and SBP, respectively) and criterion 2 (0.2 ± 2.5 and 0.9 ± 2.4 mm Hg for DBP and SBP, respectively) of the BP validation criterion of the International Organization for Standardization. Conclusion: BP measurement using a smartwatch with a photoplethysmography sensor is an accurate and reliable method in patients with PD.

Scientific assessment

01

Research question

Assess the question within the original study’s scope.

02

Methods & sample

Check the source article for study design and sample details.

03

Key findings

An editorially approved findings summary is not yet published for this record.

04

Limitations

Generalizability, bias and clinical meaning require separate assessment.

Methods, samples and clinical conclusions are not inferred from bibliographic metadata.