Hematoma-induced Twiddler-like phenomenon as a presentation of DBS hardware failure: Case report.

Front Hum Neurosci · 2023-04-21

DOI: 10.3389/fnhum.2023.1160237PMID: 37151898

Open accessJournal Article; Case Reports; case-reportOriginal source ↗

Research overview

The text below is the source article’s original abstract under a verified reuse license. cc by

Deep brain stimulators (DBS) may fail for a multitude of reasons. We present a 79-year-old Parkinson's disease patient who suffered a DBS failure following impulse generator (IPG) replacement surgery due to the IPG flipping within an expanded capsular pocket. This creation of the pocket was unintentional, and the pocket formed around an undiagnosed postoperative hemorrhage. The syndrome could be considered "Twiddler-like" because it resulted in device flipping. There were, however, many characteristic differences between our case and classical Twiddler's syndrome. DBS neurostimulator failure due to hematoma induced device flipping should be suspected when device interrogation is impossible or there are abnormally high impedances across multiple DBS lead contacts. A plain film X-ray series should be ordered and can be useful in providing radiological evidence of device flipping. In cases like ours the extensive braiding encountered in Twiddler's syndrome may be absent. Anchoring the IPG to a deep fascial layer as well as the use of an antimicrobial pouch are two methods that may be employed to prevent or to treat this complication.

Scientific assessment

01

Research question

Assess the question within the original study’s scope.

02

Methods & sample

Check the source article for study design and sample details.

03

Key findings

An editorially approved findings summary is not yet published for this record.

04

Limitations

Generalizability, bias and clinical meaning require separate assessment.

Methods, samples and clinical conclusions are not inferred from bibliographic metadata.