Analytical and clinical validity of wearable, multi-sensor technology for assessment of motor function in patients with Parkinson's disease in Japan.

Sci Rep · 2023-03-14

DOI: 10.1038/s41598-023-29382-6PMID: 36918552

Open accessResearch Support, Non-U.S. Gov't; research-article; Journal ArticleOriginal source ↗

Research overview

The text below is the source article’s original abstract under a verified reuse license. cc by

Continuous, objective monitoring of motor signs and symptoms may help improve tracking of disease progression and treatment response in Parkinson's disease (PD). This study assessed the analytical and clinical validity of multi-sensor smartwatch measurements in hospitalized and home-based settings (96 patients with PD; mean wear time 19 h/day) using a twice-daily virtual motor examination (VME) at times representing medication OFF/ON states. Digital measurement performance was better during inpatient clinical assessments for composite V-scores than single-sensor-derived features for bradykinesia (Spearman |r|= 0.63, reliability = 0.72), tremor (|r|= 0.41, reliability = 0.65), and overall motor features (|r|= 0.70, reliability = 0.67). Composite levodopa effect sizes during hospitalization were 0.51-1.44 for clinical assessments and 0.56-1.37 for VMEs. Reliability of digital measurements during home-based VMEs was 0.62-0.80 for scores derived from weekly averages and 0.24-0.66 for daily measurements. These results show that unsupervised digital measurements of motor features with wrist-worn sensors are sensitive to medication state and are reliable in naturalistic settings.Trial Registration: Japan Pharmaceutical Information Center Clinical Trials Information (JAPIC-CTI): JapicCTI-194825; Registered June 25, 2019.

Scientific assessment

01

Research question

Assess the question within the original study’s scope.

02

Methods & sample

Check the source article for study design and sample details.

03

Key findings

An editorially approved findings summary is not yet published for this record.

04

Limitations

Generalizability, bias and clinical meaning require separate assessment.

Methods, samples and clinical conclusions are not inferred from bibliographic metadata.